OF-01E Forced Convection Oven (52L)


Economy type
Stable oven that is faithful to the basic functions

  • Stable product that is faithful to the basics.
  • Includes safety device as standard.
  • 120V, 60Hz

 

OF-01E is no longer available. Please click here to view OF4-05S.

Structural Functional Features

  • Economical model series that minimizes the purchase burden.
  • Superior insulation and enclosed design minimize energy loss and reduce operating costs.
  • Stainless steel interior and shelves are excellent in terms of corrosion resistance and clean maintenance.
  • Two vent holes for gas discharge at the top.

Use Convenience Features

  • Calibration function minimizes temperature difference.
  • Optimized control with temperature auto-tuning.
  • Save and use 3 frequently used temperatures.
  • Wait On/Off timer. (up to 99 hours 59 minutes)
  • Microprocessor PID method for precise temperature control.
  • Automatic restart of operation when power is restored after sudden power failure.

Outstanding Safety

  • Top-rated overheating protection system. (registration 10-0397583)
  • Over temperature limit function.
  • Safety circuit design to protect against overcurrent and overheating.
  • Safe surface temperature even when operating at higher temperatures.
  • Malfunction prevented by controller lock function.
  • Self-diagnostics function to check abnormal condition of temperature sensor.

SKU: AAH14016U Category:
Weight 102 lbs
Dimensions 24 × 24 × 32 in
Chamber volume (L / cu ft) 52 / 1.8
Temperature
Range (℃ / ℉) Amb. +10 to 220 / Amb. +18 to 428
Fluctuation at 100℃ (±℃ / ℉) 0.4 / 0.72
Variation at 100℃ (±℃ / ℉) 3.3 / 5.94
Heating time to 100℃ (min.) 6
Recovery time at 100℃ (min.) 1
Dimension
Interior (W x D x H, mm / inch) 375 x 370 x 370 / 14.8 x 14.6 x 14.6
Exterior (W x D x H, mm / inch) 530 x 535 x 737 / 20.9 x 21.1 x 29
Net weight (Kg / lbs) 36 / 79.4
Shelves
Quantity of shelves (standard/max.) 2 / 5
Max. Load per shelf (Kg / lbs) 30 / 66.1
Electrical requirements (120V, 60Hz, A) 6.8
Cat. No. AAH14016U

※ The lowest temperature that can be controlled depends on changes in room temperature and sample temperature. Therefore, please contact the distributor of our products before purchasing for technical consultation.
※ Technical data (according to DIN 12880, before 2013)

Voltage Hertz Phase Plug Type
120 50/60 1P Type B l NEMA 5